Breath Detect

The Analyzer uses a built-in breath detection algorithm. The algorithm determines the

breath phases, rejecting signal noise and other artifacts.

Breath Detect Threshold

The Analyzer breath detection algorithm uses a flow rate threshold that the flow must

cross to trigger a change in the breath phase. Higher flow rate thresholds yield less

sensitive breath detection. Lower thresholds yield more sensitive breath detection.
To enter a new threshold, select Threshold from the Breath Detect menu under the Setup

menu, as shown in Figure A-32.

edg96s.bmp

Figure A-32. Breath Detect Threshold Selection

The dialog box shown in Figure A-33 now appears, allowing you to enter your new

value. Change the value by using the scrolling arrows or by entering a number directly.

edg73s.bmp

Figure A-33. Breath Detect Dialog Box