Oxygen Measurement

Range:

0 to 100 %

Accuracy:

± 2 % full-scale output

Resolution:

0.1 % O

2

Frequency Response:

> 15 s (t

10-90

)

Sample Rate:

100 Hz

Sensor Technology:

Galvanic Fuel Cell

Calibration:

Allows user calibration using air and 100 % O

2

Notes:

• Automatic partial pressure compensation for barometric and airway pressure changes

• Recommended interval for changing the Oxygen Sensor is one year. However, sensor may last
longer. During user calibration of the sensor, the Analyzer can detect if the sensor needs
replacement.

• Maxtec Oxygen Sensor distributed by Fluke Biomedical

Barometric Pressure Measurement

Range:

8 to 18 psia (400 to 900 mmHg)

Accuracy:

± 2 % of reading

Resolution:

0.1 mmHg

Frequency Response:

< 5 s (t

10-90

)

Sample Rate:

N/A

Calibration:

Not required; however, device allows user calibration of offset.

Temperature and Relative Humidity

Temperature Relative

Humidity

Resolution:

0.1

°

0.1 %

Range:

0 – 50

°

C

10 to 95 %

Accuracy:

± 1

°

C

± 10 % RH

Units:

°

C,

°

F,

°

K

%