Serial Communications

Selecting the COM Port

The Analyzer communicates with your PC through a serial (COM) port. With the

Analyzer connected to your computer and turned on when the software starts, the

software automatically detects which COM port the Analyzer is on and then initiates

communications between the Analyzer and your PC.
While running VT for Windows you may select Comm Port from the Communications

menu to change the port selection.

Selecting the Serial Communications Mode

The Analyzer allows for four modes for data transmission via the serial port. These are

available from the Communications menu. The currently selected mode shows a check

mark next to its name.

Both

This selection starts the automatic transmission of both waveform data and parameter
data, as described above.

Wave Form

This selection starts the continuous transmission of waveform data from the
Analyzer. The software receives flow, volume, airway pressure, auxiliary pressure,
and differential pressure data at a 50 Hz rate.

Parameters

Start the automatic transmission of parameters calculated at the end of a breath.
These parameters update on a breath-to-breath basis. Please note that no parameters
update for a breath if a zero occurs during that breath.

None

Use this selection to set no data transmission. This stops all automatic data
transmissions from the Analyzer.

Data and File Handling

File Structure

VT for Windows files handle three kinds of data: Signals, Parameters, and Events.

Signals are the raw flow, pressure, and volume time-series waveforms. Parameters are the

breath-by-breath ventilator parameters calculated by the Analyzer device. Events are

information placed into the data files by you. VT for Windows provides an interface for

you to mark events during data collection.
VT for Windows can generate three different file types:

*.vtp

The software generates .vtp files when you select to save data to a file. These files
contain signal, parameters, and event information.

*.sig

Signal file of raw waveforms. The software generates the .sig files from the .vtp files
using a file conversion program. Do this by selecting Extract Signal File from the VT
for Windows File menu. Close all files to enable this option. The .sig files are in
ASCII format; you can easily import them into your favorite spreadsheet.

*.par

Breath-by-breath calculated ventilator parameters. The software generates the .par
files from the .vtp files using a file conversion program. Do this by selecting Extract

Advertising