3. On the 682XXB/683XXB:

a. Press

MODULATION

, then

Pulse

to go to the

Internal Pulse Status Menu display.

b. Press

Edit Width

and set W1 to the pulse

width noted in the Test Record.

c. Press

On/Off

to turn on pulse modulation.

4. On the Oscilloscope, observe that the nominal

level of the displayed pulse peak has shifted ver-
tically, slightly off of the centerline. It may be
necessary to readjust the scope trigger level.

5. On the 682XXB/683XXB:

a.Press

CW/SWEEP SELECT

to return to the

CW Menu display.

b. Record the value of the output level L1, shown

in the Level display area, on the Test Record.

c. Press

Edit L1

and adjust the power level until

the nominal peak level is evenly centered on
the display centerline reference.

d. Record this value of the output level L1 on the

Test Record.

6. Subtract the L1 reading recorded in step 5.d from

the reading recorded in step 5.b. This is the pulse
level error. Record it on the Test Record.

7. On the 682XXB/683XXB:

a. Press

Edit L1

and set L1 to the maximum lev-

eled output power level for the instrument be-
ing tested (refer to Table 3-2, page 3-6).

b. Press

MODULATION

, then

Pulse

to return to

the Internal Pulse Status Menu display.

c. Press

On/Off

to turn off pulse modulation.

8. Repeat steps 1 through 7 for any other frequen-

cies and pulse widths noted in the Test Record.
Record the results on the Test Record.

3-42

682XXB/683XXB MM

PERFORMANCE

PULSE MODULATION TESTS:

VERIFICATION

RISE TIME, FALL TIME, OVERSHOOT, AND LEVEL