Anritsu 682XXB User Manual

Page 89

Advertising
background image

3. Set up the 683XXB as follows for an analog sweep

power level flatness test:

a. Reset the instrument by pressing

SYSTEM

,

then

Reset

. The CW Menu is displayed.

b. Press

Analog

to place the 683XXB in the ana-

log sweep frequency mode and display the
Analog Sweep Menu.

c. With the Analog Sweep Menu displayed, press

the main menu key

FREQUENCY

CONTROL

The Sweep Frequency Control Menu is then
displayed.

d. Press

Full

to select a full range frequency

sweep.

e. Press

Edit L1

to open the current power level

parameter for editing.

f. Set L1 to the power level noted on the test rec-

ord.

g. Now, return to the Analog Sweep Menu by

pressing the main menu key

CW/SWEEP

SELECT

h. At the Analog Sweep Menu, press the menu

soft-key

Sweep Ramp

to go to the Analog

Sweep Ramp Menu.

i. At this menu, press

Sweep Time

to open the

sweep time parameter for editing.

j. Set the sweep time to 99 seconds.

NOTE

Monitor the 683XXB’s Horizontal Output
on the Oscilloscope to determine sweep
start and stop.

4. During the analog sweep, measure the maximum

and minimum Power Meter readings and record
the values on the Test Record. Verify that the
variation (difference between the maximum and
minimum readings) does not exceed the value
noted on the Test Record.

682XXB/683XXB MM

3-29

PERFORMANCE

POWER LEVEL ACCURACY

VERIFICATION

AND FLATNESS TESTS