Step 4.

Connect a 562 Scalar Network Analyzer to the 682XXB/
683XXB as follows:

a.

Connect the 682XXB/683XXB

AUX I/O

to the 562

AUX

I/O

.

b.

Connect the 562

DEDICATED GPIB

to the 682XXB/

w683XXB

IEEE-488 GPIB

.

c.

Connect the RF Detector to the 562 Channel A Input.

Step 5.

Set up the 562 Scalar Network Analyzer as follows:

a.

Press SYSTEM MENU display.

b.

From System Menu display, select RESET.

c.

Press CHANNEL 2 DISPLAY: OFF.

d.

Press CHANNEL 1 DISPLAY: ON.

e.

Press CHANNEL 1 Menu key.

f.

From the Channel 1 Menu display, select POWER.

Step 6.

Using the scalar network analyzer, measure the RF output
at J4 of the switched filter assembly. The amplitude of the
RF signal should be >+18 dBm throughout the full sweep.

G

If the amplitude of the RF signal is correct, replace the
SQM.

G

If there is no RF signal or if the amplitude of the RF sig-
nal is low, replace the switched filter assembly.