Anritsu 682XXB User Manual

Page 287

Advertising
background image

682XXB/683XXB MM

A-49

TEST

MODEL

RECORD

68247B/68347B

3-14 Pulse Modulation Tests: Rise Time, Fall Time, Overshoot, and Level (Continued)

Pulse Level Accuracy (5 GHz, Pulse Width = 0.5

m

s)

Lower Limit

Measured Value

Upper Limit

Set F1 to 5.0 GHz
Set pulse width W1 to 5.0

m

s

Record the displayed output level L1 (step 5.b) . . . . . . . . . . . . . . . . . . . .

Record the displayed output level L1 (step 5.d) . . . . . . . . . . . . . . . . . . . .

_______ dBm

_______ dBm

Calculate and record the pulse level error. . . . . . . . . . . . . .

–1.0 dB

_______ dB

+1.0 dB

Pulse Level Accuracy (5 GHz, Pulse Width = 5.0

m

s)

Set F1 to 5.0 GHz
Set pulse width W1 to 5.0

m

s

Record the displayed output level L1 (step 5.b) . . . . . . . . . . . . . . . . . . . .

Record the displayed output level L1 (step 5.d) . . . . . . . . . . . . . . . . . . . .

_______ dBm

_______ dBm

Calculate and record the pulse level error. . . . . . . . . . . . . .

–0.5 dB

_______ dB

+0.5 dB

3-15 Pulse Modulation Test: Video Feedthrough

Test Procedure

Lower Limit

Measured Value

Upper Limit

Set F1 to 5.0 GHz
Set pulse width W1 to 5.0

m

s

Set PRI (period) to 0.01 ms

Measure and record the Video Feedthrough voltage spikes . . . .

–10 mV

_______ mV

+10 mV

Advertising