Perform the calibration as follows:

a. At the

$

prompt, type:

calterm 124

and

press <ENTER>.

b. Set up the Function Generator for a 0.4 Hz

square wave with an output level of 2 volts
peak to peak.

c. On the Spectrum Analyzer, set the Span/Div to

50 MHz per division.

d. On the computer keyboard, use the ‘, 1, 2, and

3 keys to increment and the 7, 8, 9, and 0 keys
to decrement the value of the DAC’s setting.
Start calibration by pressing the key.

e. While observing the Spectrum Analyzer dis-

play, adjust the value of the DAC’s setting to
obtain a 200 MHz peak to peak deviation. This
is the coarse adjustment.

f. On the Spectrum Analyzer, set the Span/Div to

5 MHz per division and adjust the center fre-
quency control to position the low carrier at
the center of the display. Note the frequency
reading.

g. Now, adjust the center frequency control to po-

sition the high carrier at the center of the dis-
play. Note the frequency reading.

h. The difference between these two frequencies

is the actual peak-to-peak frequency deviation.
It should be 200 MHz

±8 MHz. If not, fine ad-

just the value of the DAC’s setting to obtain
this deviation.

i. When finished setting the DAC, press

Q

on the

keyboard to go to the next calibration step (ad-
justing the DAC to obtain 20 MHz deviation).

j. Start calibration by pressing the

1

key.

k. While observing the Spectrum Analyzer dis-

play, adjust the value of the DAC’s setting to
obtain a 20 MHz peak to peak deviation. This
is the coarse adjustment.

l. On the Spectrum Analyzer, set the Span/Div to

1 MHz per division and adjust the center fre-
quency control to position the low carrier at
the center of the display. Note the frequency
reading.

682XXB/683XXB MM

4-39

FM

CALIBRATION

CALIBRATION