Test procedure 3-44, Test procedure -44 – Anritsu 682XXB User Manual

Page 104

Advertising
background image

Test
Procedure

The following procedure lets you measure the video
feedthrough that occurs during pulse modulation.

1. Set up the 682XXB/683XXB as follows:

a. Reset the instrument by pressing

SYSTEM

,

then

Reset

. Upon reset, the CW Menu is dis-

played.

b. Press

Edit F1

to open the current frequency

parameter for editing.

c. Set F1 to 5.0 GHz.

d. Press

Edit L1

to open the current power level

parameter for editing.

e. Set L1 to the maximum leveled output power

level for the instrument being tested (refer to
Table 3-2, page 3-6).

f. Press

MODULATION

, then

Pulse

. The Inter-

nal Pulse Status menu is displayed.

g. At the Internal Pulse Status menu, press

Edit Width

and set pulse width W1 to 5.0

ms.

h. Press

Edit Period

and set the PRI to 0.01 ms.

i. Press

On/Off

to turn pulse modulation on.

2. Set the Oscilloscope controls as follows:

a. Vertical Sensitivity: 5 mV/div

b. External Trigger: On the positive slope of the

Channel 1 signal input

c. Horiz Time Base: 1

ms/div

NOTE

Use the Oscilloscope’s 20 MHz bandwidth
limit to aid in viewing the voltage spikes. It
may be necessary to adjust the Oscillo-
scope’s horizontal level as any ripple or
voltage spikes are generally very small in
amplitude.

3. Measure and record the voltage spikes (video

feedthrough) on the Test Record. To meet specifi-
cations, they must be <

±10 mV.

3-44

682XXB/683XXB MM

PERFORMANCE

PULSE MODULATION TEST:

VERIFICATION

VIDEO FEEDTHROUGH