Alc slope calibration calibration – Anritsu 682XXB User Manual

Page 139

Advertising
background image

d. Press

CW/SWEEP SELECT

to return to the

Step Sweep Menu display.

e. Press

Sweep Ramp

. At the resulting Step

Sweep Ramp Menu, press

Num of Steps

and

set the number of steps to 400.

3. Make the following selections on the 562 Network

Analyzer to normalize the step sweep.

a. Press

CALIBRATION

and follow the menu on

the display.

b. Press

AUTOSCALE

.

c. Press

OFFSET/RESOLUTION

and set the Reso-

lution to 0.5 dB.

4. On the 683XXB, press

Analog

to select the ana-

log sweep mode.

5. Adjust the ALC Slope as follows:

a. At the

$

prompt on the PC display, type:

slpcal

and press <ENTER>.

On the computer keyboard, the adjustment
keys are:

Slope

E (Up)

D (Down)

(all bands)

Offset

Q (Up)

A (Down)

(band 1-4 only)

682XXB/683XXB MM

4-29

ALC SLOPE

CALIBRATION

CALIBRATION

8.4 GHz

20 GHz

40 GHz

ALC Slope Before Adjustment

Normalized

Step Sweep

50. 60, or

65 (67) GHz

2 GHz

Figure 4-8.

ALC Slope Adjustment Waveform Display