5-46

682XXB/683XXB MM

Step 5.

Set up the 562 Scalar Network Analyzer as follows:

a.

Press SYSTEM MENU key.

b.

From System Menu display, select RESET.

c.

Press CHANNEL 2 DISPLAY: OFF.

d.

Press CHANNEL 1 DISPLAY: ON.

e.

Press CHANNEL 1 MENU key.

f.

From the Channel 1 Menu display, select POWER.

Step 6.

Using the scalar network analyzer, measure the RF output
at J2 of the switched filter assembly. The amplitude of the
RF signal should be >+15 dbm (>+20 dBm with Option 15B)
throughout the full sweep.

G

If the amplitude of the RF signal is correct, check for bad
cables.

G

If there is no RF signal or if the amplitude of the RF sig-
nal is low, replace the switched filter assembly.

Table 5-20.

Error Messages 128, 129, 130, 131, 132, 133, 134, and 135 (6 of 8)

TROUBLESHOOTING

TROUBLESHOOTING

TABLES