Anritsu 682XXB User Manual

Page 116

Advertising
background image

4-6

682XXB/683XXB MM

Subassembly/RF Component Replaced

Perform the Following

Calibration(s) in Paragraph(s):

A1, A2 Front Panel Assy

None

A3 Reference Loop PCB

4-7

A4 Coarse Loop PCB

4-7

A5 Fine Loop PCB

4-7

A6 Pulse Generator PCB

None

A7 YIG Loop PCB

None

A8 Function Generator PCB

4-12

A9 PIN Control PCB

4-8, 4-9, 4-10, 4-11

A10 ALC PCB

4-8, 4-9, 4-10, 4-11, 4-12

A11 FM PCB

4-13 (and 4-14 if Option 6 is installed)

A12 Analog Instruction PCB

4-7

A13 YIG Driver PCB

4-7

A14 SDM, SQM Driver PCB

4-9, 4-10, 4-11

A15 Regulator PCB

None

A16 CPU Interface PCB

4-7

A17 CPU PCB

4-7 thru 4-13. None, if calibration EEPROM reused.

A18 Power Supply PCB

None

A19 AC Line Conditioner PCB

None

A21 Line Filter/Rectifier PCB

None

A21-1 BNC/Aux IO Connector PCB

None

YIG-tuned Oscillator

4-7

0.01 to 2 GHz Down Converter Assy

4-8, 4-9, 4-10, 4-11

0.5 to 2.2 GHz Digital Down Converter Assy

4-8, 4-9, 4-10, 4-11

Switched Filter Assy

4-8, 4-9, 4-10, 4-11

Switched Doubler Module (SDM)

4-8, 4-9, 4-10, 4-11

Source Quadrupler Module (SQM)

4-8, 4-9, 4-10, 4-11

Forward Coupler

4-8, 4-9, 4-10, 4-11

Directional Coupler

4-8, 4-9, 4-10, 4-11

Step Attenuator

4-8, 4-9, 4-10

Table 4-2.

Calibration Following Subassembly/RF Component Replacement

CALIBRATION FOLLOWING

CALIBRATION

SUBASSEMBLY REPLACEMENT

Advertising
This manual is related to the following products: