Unleveled with no/low output power:

Step 1.

Set up the 682XXB/683XXB as follows:

a.

682XXB Setup:
CW/SWEEP SELECT: Step
F1: 0.010 GHz
F2: 20.000 GHz
Number of Steps: 400
L1: +1.00 dBm

683XXB Setup:
CW/SWEEP SELECT: Analog
F1: 0.010 GHz
F2: 20.000 GHz
Sweep Time: 0.100 Sec
L1: +1.00 dBm

b.

LEVEL/ALC SELECT: ALC Mode
Leveling Menu: Internal

Step 2.

Connect the X input of an oscilloscope to the 682XXB/
683XXB rear panel

HORIZ OUT

connector.

Step 3.

Using the oscilloscope, check the switched filter bias volt-
ages at A13TP4 and A13TP9. The bias voltage at A13TP4
should be +6 volts; the bias voltage at A13TP9 should be +8
volts. If the 682XXB/683XXB has a SDM installed, also
check for a +20 volt PIN switch drive voltage at A9TP9.

G

If the bias and the PIN switch drive voltages are correct,
go to step 4.

G

If the bias voltages are not correct, replace the A13 PCB.

G

If the PIN switch drive voltage is not correct, replace the
A9 PCB.

Step 4.

Connect a 562 Scalar Network Analyzer to the 682XXB/
683XXB as follows:

a.

Connect the 682XXB/683XXB

AUX I/O

to the 562

AUX

I/O

.

b.

Connect the 562

DEDICATED GPIB

to the 682XXB/

683XXB

IEEE-488 GPIB

.

c.

Connect the RF Detector to the 562 Channel A Input.

Table 5-20.

Error Messages 128, 129, 130, 131, 132, 133, 134, and 135 (5 of 8)