Anritsu 682XXB User Manual

Page 363

Advertising
background image

682XXB/683XXB MM

A-125/A-126

TEST

MODEL

RECORD

68263B/68363B

3-15 Pulse Modulation Test: Video Feedthrough

Test Procedure

Lower Limit

Measured Value

Upper Limit

Set F1 to 5.0 GHz
Set pulse width W1 to 5.0

m

s

Set PRI (period) to 0.01 ms

Measure and record the Video Feedthrough voltage spikes . . . .

–10 mV

_______ mV

+10 mV

3-16 Pulse Modulation Test: RF On/Off Ratio

Test Procedure

Lower Limit

Measured Value

Upper Limit

Set F1 to 5.0 GHz

Measure and record the peak of the signal on the Spectrum
Analyzer. (Measured signal must be >60 dB below top graticule
to meet specification; this represents an On/Off Ratio of >80 dB.) .

60 dB

_______ dB

N/A

3-17 Phase Modulation Tests

F

M Imput Sensitivity Procedure

(Wide

F

M Mode)

Lower Limit

Measured Value

Upper Limit

Set F1 to 5.0 GHz

Record the

F

M deviation displayed on the 68263B/68363B . . . . . . . . . . . . . .

_______rad

Calculate and record the

F

M accuracy . . . . . . . . . . . . . . .

93.7%

_________ %

106.3%

F

M Imput Sensitivity Procedure

(Narrow

F

M Mode)

Set F1 to 5.0 GHz

Record the

F

M deviation displayed on the 68263B/68363B . . . . . . . . . . . . . .

_______rad

Calculate and record the

F

M accuracy . . . . . . . . . . . . . . .

93.7%

_________ %

106.3%

Advertising