Calibration steps 4-14, Calibration steps -14, Preliminary calibration calibration – Anritsu 682XXB User Manual

Page 124

Advertising
background image

Calibration
Steps

Each of the steps in this procedure provides initial
calibration of a specific 682XXB/683XXB circuit or
component. To ensure accurate instrument calibra-
tion, each step of this procedure must be performed
in sequence.

1. Calibrate the internal DVM circuitry as follows:

a. At the

$

prompt, type:

calterm 119

and

press <ENTER>. (The

$

prompt will appear on

the screen when the calibration is complete.)

b. Record step completion on the Test Record.

2. Calibrate the Coarse Loop Pretune DAC as fol-

lows:

a. At the

$

prompt, type:

calterm 137

and

press <ENTER>. (The

$

prompt will appear on

the screen when the calibration is complete.)

b. Record step completion on the Test Record.

3. Calibrate the Fine Loop Pretune DAC as follows:

a. At the

$

prompt, type:

calterm 136

and

press <ENTER>. (The

$

prompt will appear on

the screen when the calibration is complete.)

b. Record step completion on the Test Record.

4. Calibrate the Sweep Time DAC as follows:

a. At the

$

prompt, type:

calterm 132

and

press <ENTER>. (The

$

prompt will appear on

the screen when the calibration is complete.)

b. Record step completion on the Test Record.

5. Calibrate the YIG Frequency Offset DAC as fol-

lows:

a. At the

$

prompt, type:

calterm 134

and

press <ENTER>.

b. Follow the instructions on the screen.

NOTE

Adjust the DAC to the number that will get
the closest to the 1.9 GHz and 8.3 GHz
called for by the procedure.

c. Record step completion on the Test Record.

4-14

682XXB/683XXB MM

PRELIMINARY

CALIBRATION

CALIBRATION

NOTE

To save the calibration data after
completing any calibration step,
type:

calterm 787

and press

<ENTER>.