Troubleshooting, Troubleshooting tables – Anritsu 682XXB User Manual

Page 210

Advertising
background image

5-44

682XXB/683XXB MM

Error 129 Switched Filter or Level Detector Failed

Description: Error 129 indicates a failure of either the switched filter
or level detector circuitry. The 682XXB/683XXB may or may not pro-
duce an RF output in the 2 to 20 GHz frequency range. Thus, there are
two troubleshooting paths for this problem—unleveled with output
power and unleveled with no/low output power.

Unleveled with output power (The warning message

UNLEVELED

appears on the front panel display):

Step 1.

Set up the 682XXB/683XXB as follows:

a.

682XXB Setup:
CW/SWEEP SELECT: Step
F1: 0.010 GHz
F2: 20.000 GHz
Number of Steps: 400
L1: +1.00 dBm

683XXB Setup:
CW/SWEEP SELECT: Analog
F1: 0.010 GHz
F2: 20.000 GHz
Sweep Time: 0.100 Sec
L1: +1.00 dBm

b.

LEVEL/ALC SELECT: ALC Mode
ALC Mode: Ext ALC Front
Leveling Menu: External Detector

Step 2.

Connect a detector to the 682XXB/683XXB

RF OUTPUT

con-

nector and connect the detected DC output of the detector to
the front panel

EXTERNAL ALC IN

connector.

G

If the warning message

UNLEVELED

no longer appears

on the front panel display, replace the directional coupler.

G

If the warning message

UNLEVELED

is still displayed,

replace the A10 PCB.

Table 5-20.

Error Messages 128, 129, 130, 131, 132, 133, 134, and 135 (4 of 8)

TROUBLESHOOTING

TROUBLESHOOTING

TABLES