11 power level accuracy and flatness tests 3-26, Test setup 3-26, Power level accuracy and flatness tests, 3-26 – Anritsu 682XXB User Manual

Page 86: Power level accuracy and flatness, Tests -26, Test setup -26

Advertising
background image

3-11

POWER LEVEL
ACCURACY AND
FLATNESS TESTS

The following tests can be used to verify that the 682XXB/683XXB
meets its power level specifications. Power level verification testing is
divided into two parts—power level accuracy tests and power level
flatness tests.

Test Setup

Connect the equipment, shown in Figure 3-6, as fol-
lows:

1. Calibrate the Power Meter with the Power Sen-

sor.

2. Connect the Power Sensor to the

RF OUTPUT

of

the 682XXB/683XXB.

NOTE

For

£40 GHz models, use the K (male) to

2.4 mm (female) adapter to connect the
Power Sensor to the

RF OUTPUT

connec-

tor.

3. Connect the 682XXB/683XXB rear panel

HORIZ

OUT

to the Oscilloscope CH.1 input (X input).

NOTE

During this test it will be necessary to
adjust the Power Meter’s CAL FACTOR %
setting as applicable for the frequency
being tested.

3-26

682XXB/683XXB MM

RF

OUT

HORIZ

OUT

682XXB / 683XXB

OSCILLOSCOPE

Power

Sensor

POWER METER

CH 1 or X Input

Figure 3-6.

Equipment Setup for Power Level Accuracy and Flatness Tests

PERFORMANCE

POWER LEVEL ACCURACY

VERIFICATION

AND FLATNESS TESTS