Switched filter calibration shaper calibration – Anritsu 682XXB User Manual

Page 131

Advertising
background image

Shaper DAC
Adjustment

For Models 682XXB with Firmware Version
2.39 and above; Models 683XXB with Firmware
Version 2.46 and above; Models 682X5B with
Firmware Version 1.28 and above; and Models
683X5B with Firmware Version 1.32 and above.

The following step in the procedure adjusts the
Switch Filter Shaper DAC which controls the gain of
the Switched Filter Shaper Amplifier. Each fre-
quency band will be scanned for the minimum un-
leveled power point before automatic adjustment of
the shaper DAC.

1. At the

$

prompt on the PC display, type:

calterm 138

and press <ENTER>. (The

$

prompt will appear on the screen when the cali-
bration is complete.)

NOTE

The calibration routine may take up to 20
minutes depending on the frequency range
of the 682XXB/683XXB being calibrated.

2. Store the calibration data in non-volatile memory

(EEPROMs) on the A17 CPU PCB as follows:

a. Type:

calterm 787

and press <ENTER>.

(The

$

prompt will appear on the screen when

the data has been stored.)

b. Record step completion on the Test Record.

682XXB/683XXB MM

4-21

SWITCHED FILTER

CALIBRATION

SHAPER CALIBRATION

CAUTION

When saving calibration data, turn-
ing off the instrument before the

$

prompt returns to the screen can
cause all stored data to be lost.