NOTE

Before beginning this calibration proce-
dure, always let the 682XXB/683XXB
warm up for a minimum of one hour.

ALC Slope
DAC
Adjustment

The following procedure lets you adjust the ALC
Slope over individual frequency ranges to compen-
sate for decreasing output power-vs-frequency in
analog sweep.

The procedure begins by letting you adjust the ALC
Slope for band 0 (0.01 to 2.0 GHz), if installed. It
then continues letting you adjust the ALC Slope
from 2 GHz to the top frequency of the instrument
in up to four bands. The band frequency ranges are:

Band 1

2.0 to 8.4 GHz

Band 2

8.4 to 20 GHz

Band 3

20.0 to 40.0 GHz

Band 4

40.0 to 65.0 GHz

During band 1 thru 3/4 ALC Slope adjustment, the
562 Network Analyzer display (Figure 4-8) shows
the response from 2 GHz to the top frequency of the
model, as adjustment is done band by band.

1. Set up the 562 Network Analyzer as follows:

a. Press the System Menu key.

b. From the System Menu display, select

RESET

.

c. Press

CHANNEL 2 DISPLAY: OFF

.

d. Press

CHANNEL 1 DISPLAY: ON

.

e. Press

CHANNEL 1 MENU

key.

f. From the Chanel 1 Menu Display, select

TRANSMISSION

and

SELECT INPUT

(NON-RATIO A)

.

2. Set up the 683XXB as follows:

a. Reset the instrument by pressing

SYSTEM

then

Reset

. Upon reset the CW Menu is dis-

played.

b. Press

Step

. The Step Sweep Menu is dis-

played.

c. Press

FREQUENCY CONTROL

, then

Full

to

select the full frequency range of the unit being
calibrated.

4-28

682XXB/683XXB MM

ALC SLOPE

CALIBRATION

CALIBRATION

NOTE

For 683X5B models, the frequency
ranges for Band 0 and Band 1 are
0.5 to 2.2 GHz and 2.2 to 8.4 GHz re-
spectively.