Anritsu 682XXB User Manual

Page 97

Advertising
background image

6. Compute the FM accuracy with the following for-

mula:

Accuracy in

FM Deviation in kHz

(

%)

(

)

=

ж
и

з

ц
ш

ч ´

240

100

7. The calculated result should be between 93.7%

and 106.3%. Record this result on the Test Re-
cord.

Locked Low-Noise FM Mode

1. Set up the 682XXB/683XXB as follows:

a. Reset the instrument by pressing

SYSTEM

,

then

Reset

. Upon reset, the CW Menu is dis-

played.

b. Press

Edit F1

to open the current frequency

parameter for editing.

c. Set F1 to 5.0 GHz.

2. Set up the Spectrum Analyzer as follows:

a. CF: 5.0 GHz

b. Span/Div: 100 kHz (1 MHz across the display)

3. On the 682XXB/683XXB, make the following set-

tings:

a. Press

MODULATION

, then

FM

. At the result-

ing FM Status menu, press

More

to go to the

additional FM Status menu.

b. At the additional FM Status menu, press

Locked LN

, then press

Previous Menu

to

return to the main FM Status menu.

c. At the FM Status menu, press

Edit Dev.

and

set the deviation to 0.240 MHz.

d. Press

Edit Rate

and set the rate to 99.8 kHz.

e. Press

On/Off

to FM on.

4. On the 682XXB/683XXB, use the cursor control

key to place the cursor under the last 0 in the de-
viation setting (0.240 MHz). While observing the
first Bessel null (Figure 3-9), increase/decrease
the deviation frequency with the cursor control
key to reduce the carrier as low as possible.

5. Record the FM deviation frequency, displayed on

the 682XXB/683XXB, on the Test Record.

682XXB/683XXB MM

3-37

PERFORMANCE

FREQUENCY

VERIFICATION

MODULATION TESTS

Advertising