Am calibration calibration – Anritsu 682XXB User Manual

Page 146

Advertising
background image

5. Perform AM Function Generator calibration as

follows:

a. At the

$

prompt, type:

calterm 146

and

press <ENTER>. (The

$

prompt will appear on

the screen when the calibration is complete.)

b. Record step completion on the Test Record.

6. Store the calibration data as follows:

a. At the

$

prompt, type:

calterm 787

and

press <ENTER>. (The

$

prompt will appear on

the screen when the calibration data has been
stored.)

b. Record step completion on the Test Record.

4-36

682XXB/683XXB MM

AM

CALIBRATION

CALIBRATION

CAUTION

When saving calibration data, turn-
ing off the instrument before the

$

prompt returns to the screen can
cause all stored data to be lost.