Troubleshooting, Troubleshooting tables – Anritsu 682XXB User Manual

Page 200

Advertising
background image

5-34

682XXB/683XXB MM

Step 8.

Run self-test.

G

If error 113 or 115 is not displayed, the problem is cleared.

G

If either error 113 or 115 is displayed, contact your local

ANRITSU

service center for assistance.

Table 5-13.

Error Messages 113 and 115 (2 of 2)

TROUBLESHOOTING

TROUBLESHOOTING

TABLES

Down Converter

Error 114 Down Converter LO not Locked

Description: The local oscillator in the down converter assembly is
not phase-locked.

Step 1.

Disconnect cable W112 at A3J2.

Step 2.

Using a spectrum analyzer, verify the presence of a +5 dBm
±6 dB, 500 MHz signal at A3J2.

G

If present, go to step 3.

G

If not present, replace the A3 PCB.

Step 3.

Reconnect cable W112 to A3J1, then disconnect cable W112
at J2 of the down converter assembly.

Step 4.

Using a spectrum analyzer, verify the presence of a +5 dBm
±6 dB, 500 MHz signal at the end of cable W112.

G

If present, replace the down converter assembly.

G

If not present, replace cable W112.

Table 5-14.

Error Message 114