Troubleshooting, Troubleshooting tables – Anritsu 682XXB User Manual

Page 217

Advertising
background image

682XXB/683XXB MM

5-51

Error 144 RF was Off when Selftest started. Some tests where
not performed

Description: Indicates that some self-tests where not performed be-
cause the RF Output was selected

OFF

on the front panel.

Step 1.

Press the

OUTPUT

key on the front panel to turn the RF

Output

ON

.

Step 2.

Run self-test again.

Table 5-24.

Error Message 144

TROUBLESHOOTING

TROUBLESHOOTING

TABLES

Error 145 AM Meter or associated circuitry failed

Description: Indicates a failure of the internal amplitude modulation
function. The 682XXB/683XXB may or may not provide amplitude
modulation of the RF output signal using modulating signals from an
external source.

Step 1.

Set up the 682XXB/683XXB as follows:

a.

Press

SYSTEM

, then

Reset

.

b.

Press

MODULATION

, then

AM

to go to the Internal AM

Status display.

c.

Press

On/Off

to turn internal amplitude modulation on.

Step 2.

Using an oscilloscope, verify the presence of a 10 volt peak to
peak sine wave signal with a period of 100

ms at the rear

panel

AM OUT

connector.

G

If present, replace the A10 PCB.

G

If not present, replace the A8 PCB.

Table 5-25.

Error Message 145