Fm calibration calibration – Anritsu 682XXB User Manual

Page 152

Advertising
background image

quency control to position the high (or low) car-
rier at the center of the display. Note the
frequency reading.

j. On the keyboard, enter

1

to toggle the polarity.

k. Now, adjust the center frequency control to po-

sition the low (or high) carrier at the center of
the display. Note the frequency reading.

l. The difference between these two frequencies

should be 200 MHz

±8 MHz. If not, enter

2

and fine adjust the DAC to obtain this devia-
tion.

m.When finished setting the DAC, press

Q

on the

keyboard to return to the menu then enter

3

to

go to the next calibration step (adjusting the
DAC to obtain 20 MHz deviation).

n. On the keyboard, enter

2

, then press the

1

key

to start calibration.

o. While oberving the Spectrum Analyzer display,

adjust the value of the DAC’s setting to obtain
a 10 MHz change in frequency.

p. On the keyboard, press

Q

to return to the

menu, then enter

1

to toggle the polarity.

q. Measure the change in frequency on the Spec-

trum Analyzer. The change in frequency
should be 20 MHz. If not, enter

2

and adjust

the DAC.

r. Repeat steps 5.p and 5.q as necessary to obtain

a 20 MHz frequency change. When finished ad-
justing the DAC, press

Q

to return to the

menu.

s. On the Spectrum Analyzer, set the Span/Div to

1 MHz per division and adjust the center fre-
quency control to position the high (or low) car-
rier at the center of the display. Note the
frequency reading.

t. On the keyboard, enter

1

to toggle the polarity.

u. Now, adjust the center frequency control to po-

sition the low (or high) carrier at the center of
the display. Note the frequency reading.

v. The difference between these two frequencies

should be 20 MHz

±0.8 MHz. If not, enter

2

and fine adjust the DAC to obtain this devia-
tion.