Phase modulation calibration ( f m) calibration – Anritsu 682XXB User Manual

Page 166

Advertising
background image

c. On the Spectrum Analyzer, set the Span/Div ot

50 kHz per division.

d. On the computer keyboard, use the ‘, 1, 2, and

3 keys to increment and the 7, 8, 9, and 0 keys
to decrement the value of the DAC’s setting.
Start the calibration by pressing an increment
key.

e. While observing the first Bessel null (Figure

4-16) on the Spectrum Analyzer display, adjust
the value of the DAC’s setting to reduce the
carrier level as low as possible.

f. When finished setting the DAC, press

Q

on the

keyboard.

The program will exit to the

$

prompt.

g. Record step completion on the Test Record.

7. Store the calibration data as follows:

a. At the

$

prompt, type:

calterm 787

and

press <ENTER>. (The

$

prompt will appear on

the screen when the calibration data has been
stored.)

b. Record step completion on the Test Record.

4-56

682XXB/683XXB MM

PHASE MODULATION

CALIBRATION

(

FM) CALIBRATION

NOTE

You may need to adjust the RBW
setting on the Spectrum Analyzer in
order to see the >–40 dBc null.

CAUTION

When saving calibration data, turn-
ing off the instrument before the

$

prompt returns to the screen can
cause all stored data to be lost.