6. Calibrate the YIG Frequency Linearizer DACs as

follows:

a. At the

$

prompt, type:

calterm 127

and

press <ENTER>.

b. Follow the instructions on the screen. Enter

the value of the frequency counter reading as
XXXX MHz.

c Record step completion on the Test Record.

7. Calibrate the 100 MHz Reference Oscillator as

follows:

a. If Option 16 (High Stability Time Base) is in-

stalled, disconnect the cable at A3J6.

b. Connect the frequency counter to the 682XXB/

683XXB RF output connector.

c. At the

$

prompt, type:

calterm 130

and

press <ENTER>.

d. Follow the instructions on the screen.

e. Reconnect the cable to A3J6, if removed.

f. If Option 16 is installed, use a Phillips screw-

driver and remove the screw on top of the
10 MHz High Stability Crystal Oscillator as-
sembly to gain access to the timebase adjust-
ment screw.

g. Using the Phillips screwdriver, adjust the ti-

mebase to obtain a frequency counter reading
of exactly 10 GHz.

h. Record step completion on the Test Record.

8. Calibrate the Ramp Center DAC as follows:

a. At the

$

prompt, type:

calterm 129

and

press <ENTER>. (The

$

prompt will appear on

the screen when the calibration is complete.)

b. Record step completion on the Test Record.

9. Calibrate the Sweep Width DAC as follows:

a. At the

$

prompt, type:

calterm 133

and

press <ENTER>. (This calibration can take ap-
proximately 2 minutes to complete.)

The

$

prompt will appear on the screen when

the calibration is complete.

b. Record step completion on the Test Record.

682XXB/683XXB MM

4-15

PRELIMINARY

CALIBRATION

CALIBRATION

NOTE

Because the 100 MHz Calibration
DAC is an 8-bit DAC, adjustment
resolution is typically 400 Hz/step.