FM CALIBRATION

Models 682XXB/683XXB with Firmware Versions 1.01 to 2.10
Models 682X5B/683X5B with Firmwave Version 1.00
This procedure provides the steps necessary to perform FM calibra-
tion. This consists of calibrating the FM Meter circuit, the Offset Bal-
ance DAC, and the FM Gain Control DAC. The FM Gain Control DAC
is calibrated for input sensitivities in both narrow and wide FM
modes.

Equipment
Setup

Connect the equipment, shown in Figure 4-11, as
follows:

1. Interface the PC to the 682XXB/683XXB by per-

forming the initial setup procedure, pages 4-7 to
4-12.

2. Connect the 682XXB/683XXB rear panel

10 MHz

REF OUT

to the Spectrum Analyzer External Ref-

erence input.

3. Connect the Function Generator Output to the

BNC tee. Connect one leg of the tee to the
682XXB/683XXB front panel

FM IN

. Connect the

other leg of the tee to the DMM input.

682XXB/683XXB MM

4-37

FM

CALIBRATION

CALIBRATION

Serial

I/O

682XXB / 683XXB

COM1

or

COM2

IBM-Compatible PC

Function

Generator

10 MHz

REF OUT

EXT REF

INPUT

RF

IN

Spectrum

Analyzer

OUTPUT

FM

IN

RF

OUTPUT

BNC
Tee

INPUT

DMM

Figure 4-11.

Equipment Setup for FM Calibration