A-11

3-11 Power Level Accuracy And Flatness Tests (Continued)

(Model 68237B/68337B with Option 15B High Power & Option 2A Step Attenuator)

Power Level Accuracy Test Procedure

Set F1 to 5.0 GHz

Set L1 to:

Measured Power *

+15 dBm

+14 dBm

+13 dBm

+12 dBm

+11 dBm

+10 dBm

+ 9 dBm

+ 8 dBm

+ 7 dBm

+ 6 dBm

+ 5 dBm

+ 4 dBm

+ 3 dBm

_________dBm

_________dBm

_________dBm

_________dBm

_________dBm

_________dBm

_________dBm

_________dBm

_________dBm

_________dBm

_________dBm

_________dBm

_________dBm

* Specification is

±

1.0 dB.

Power Level Flatness Test Procedure (Step Sweep)

Set L1 to:

Max Power

Min Power

Variation **

+15 dBm

_________dBm

__________dBm

_________dB

** Maximum variation is 1.6 dB.

Power Level Flatness Test Procedure (Analog Sweep) (Model 68337B only)

Set L1 to:

Max Power

Min Power

Variation ***

+15 dBm

_________dBm

__________dBm

_________dB

*** Maximum variation is 6.0 dB (typical, not a specification).