Unlocked Wide FM Mode

1. Set up the 682XXB/683XXB as follows:

a. Reset the instrument by pressing

SYSTEM

,

then

Reset

. Upon reset, the CW Menu is dis-

played.

b. Press

Edit F1

to open the current frequency

parameter for editing.

c. Set F1 to 5.0 GHz.

2. Set up the Spectrum Analyzer as follows:

a. CF: 5.0 GHz

b. Span/Div: 50 MHz (500 MHz across the dis-

play)

3. On the 682XXB/683XXB, make the following set-

tings:

a. Press

MODULATION

, then

FM

. At the result-

ing FM Status menu, press

More

to go to the

additional FM Status menu.

b. At the additional FM Status menu, press

Unlocked Wide

, then press

Previous Menu

to

return to the main FM Status menu.

c. At the FM Status menu, press

Mod. Wave

to

go to the Modulation Waveform Selection
menu. Use the cursor control key to highlight
the squarewave modulation waveform, then
press

Select

.

d. Press

Previous Menu

to return to the FM

Status menu, then press

Edit Dev.

and set the

deviation to 100 MHz. Press

Edit Rate

and set

the rate to 0.4 Hz.

e. Press

On/Off

to turn FM on.

4. On the Spectrum Analyzer, change the Span/Div

setting to 5 MHz to use the analyzer as a fre-
quency counter.

a. Adjust the center frequency control to position

the low carrier at the center of the display. Re-
cord the frequency reading on the Test Record.

b. Adjust the center frequency control to position

the high carrier at the center of the display.
Record the frequency reading on the Test Re-
cord.