Alc slope calibration calibration – Anritsu 682XXB User Manual

Page 136

Advertising
background image

f. Press

Sweep Ramp

. At the resulting Step

Sweep Ramp Menu, press

Num of Steps

and

set the number of steps to 200.

g. Press

Step

to return to the Step Sweep Menu

display.

7. Make the following selections on the 562 Network

Analyzer to normalize the step sweep.

a. Press CALIBRATION and follow the menu on

the display.

b. Press AUTOSCALE.

c. Press OFFSET/RESOLUTION and set the

Resolution to 0.2 dB.

8. On the 683XXB, press

Analog

to select the ana-

log sweep mode.

9. Adjust the ALC Slope DAC for frequencies

>2 GHz as follows:

a. At the

$

prompt on the PC display, type:

calset

and press <ENTER>.

b. Type

1

and press <ENTER>.

c. Select item 11 (ALC Slope Main) from the

menu. Follow the instructions presented at the
bottom of the screen to enter a new DAC set-
ting value.

d. Enter different DAC setting values to find the

setting that adjusts the slope so that the power
at the start and stop frequencies match the
normalized straight line in step sweep mode.

e. When finished, exit the program.

f. Type

X

and press <ENTER> to exit the calset

menu. (The

$

prompt will appear on the

screen.)

g. Record step completion on the Test Record.

10. Store the new DAC setting values in non-volatile

memory (EEPROMs) on the A17 CPU PCB as fol-
lows:

a. Type:

calterm 787

and press <ENTER>.

(The

$

prompt will appear on the screen when

the data has been stored.)

b. Record step completion on the Test Record.

4-26

682XXB/683XXB MM

ALC SLOPE

CALIBRATION

CALIBRATION

CAUTION

When saving calibration data, turn-
ing off the instrument before the

$

prompt returns to the screen can
cause all stored data to be lost.