Alc bandwidth calibration calibration – Anritsu 682XXB User Manual

Page 142

Advertising
background image

The

$

prompt will appear on the screen when

the ALC Bandwidth calibration is complete.
(This can take up to 15 minutes depending on
the frequency range of the 682XXB/683XXB.)

b. Record step completion on the Test Record.

2. Store the calibration data as follows:

a. At the

$

prompt, type:

calterm 787

and

press <ENTER>. (The

$

prompt will appear on

the screen when the calibration data has been
stored.)

b. Record step completion on the Test Record.

4-32

682XXB/683XXB MM

CAUTION

When saving calibration data, turn-
ing off the instrument before the

$

prompt returns to the screen can
cause all stored data to be lost.

ALC BANDWIDTH

CALIBRATION

CALIBRATION