5. The difference between these two frequencies

is the actual peak-to-peak frequency deviation.
It should be between 190 MHz and 210 MHz.
Record the difference on the Test Record.

Unlocked Narrow FM Mode

1. Set up the 682XXB/683XXB as follows:

a. Reset the instrument by pressing

SYSTEM

,

then

Reset

. Upon reset, the CW Menu is dis-

played.

b. Press

Edit F1

to open the current frequency

parameter for editing.

c. Set F1 to 5.0 GHz.

2. Set up the Spectrum Analyzer as follows:

a. CF: 5.0 GHz

b. Span/Div: 100 kHz (1 MHz across the display)

3. On the 682XXB/683XXB, make the following set-

tings:

a. Press

MODULATION

, then

FM

. At the result-

ing FM Status menu, press

More

to go to the

additional FM Status menu.

b. At the additional FM Status menu, press

Unlocked Narrow

, then press

Previous Menu

to return to the main FM Status menu.

c. At the FM Status menu, press

Edit Dev.

and

set the deviation to 0.240 MHz.

d. Press

Edit Rate

and set the rate to 99.8 kHz.

e. Press

On/Off

to FM on.

4. Since the 682XXB/683XXB is now in Unlocked

Narrow FM mode, it is necessary to retune the
Spectrum Analyzer to center the display.

5. On the 682XXB/683XXB, use the cursor control

key to place the cursor under the last 0 in the de-
viation setting (0.240 MHz). While observing the
first Bessel null (Figure 3-9), increase/decrease
the deviation frequency with the cursor control
key to reduce the carrier as low as possible.

6. Record the FM deviation frequency, displayed on

the 682XXB/683XXB, on the Test Record.

682XXB/683XXB MM

3-35

PERFORMANCE

FREQUENCY

VERIFICATION

MODULATION TESTS

CENTER FREQUENCY

BESSEL

NULL

Figure 3-9.

Typical Spectrum Analyzer Display of

Bessel Null on FM Waveform

NOTE

You may need to adjust the RBW
setting on the Spectrum Analyzer in
order to see the >–48 dBc null.