Fm calibration calibration – Anritsu 682XXB User Manual

Page 157

Advertising
background image

4. Internal Unlocked Wide FM Mode Sensitivity

calibration is accomplished by adjusting the FM
Variable Gain Control DAC to obtain a 200 MHz
FM deviation at frequencies of 5 GHz and
15 GHz. The modulating signal input is from the
internal Function Generator.

Perform the calibration as follows:

a. At the

$

prompt, type:

calterm 143

and

press <ENTER>.

The screen will display the following menu:
1) Toggle polarity (this lets you go from a

positive deviation to negative deviation.)

2) Change DAC value.
3) Quit.

b. On the Spectrum Analyzer, set the Span/Div to

50 MHz per division.

c. On the computer keyboard, enter

2

to begin

calibration.

d. On the computer keyboard, use the ‘, 1, 2, and

3 keys to increment and the 7, 8, 9, and 0 keys
to decrement the value of the DAC’s setting.
Start calibration by pressing the key.

e. While observing the Spectrum Analyzer dis-

play, adjust the value of the DAC’s setting to
obtain a 100 MHz change in frequency.

f. On the keyboard, press

Q

to return to the

menu, then enter

1

to toggle the polarity.

g. Measure the change in frequency on the Spec-

trum Analyzer. The change in frequency
should be 200 MHz. If not, enter

2

and adjust

the DAC.

h. Repeat steps 4.f and 4.g as necessary to obtain

a 200 MHz frequency change. When finished
adjusting the DAC, press

Q

to return to the

menu.

i. On the Spectrum Analyzer, set the Span/Div to

5 MHz per division and adjust the center fre-
quency control to position the high (or low)
carrier at the center of the display. Note the
frequency reading.

j. On the keyboard, enter

1

to toggle the polarity.

682XXB/683XXB MM

4-47

FM

CALIBRATION

CALIBRATION