Test procedure (2 to 10 ghz) 3-19, Test procedure (2 to 10 ghz) -19 – Anritsu 682XXB User Manual

Page 79

Advertising
background image

Test
Procedure
(2 to 10 GHz)

The following procedure lets you measure the 2 to
10 GHz (2.2 to 10 GHz for 68XX5Bs) RF output har-
monic levels to verify that they meet specifications.

1. Set up the 682XXB/683XXB as follows:

a. Reset the instrument by pressing

SYSTEM

,

then

Reset

. Upon reset the CW Menu is dis-

played.

b. Press

Edit L1

to open the current power level

parameter for editing.

c. Set L1 to the lesser of +10 dBm or the maxi-

mum leveled power level for the instrument
being tested (refer to Table 3-2, page 3-6).

d. Press

Edit F1

to open the current frequency

parameter for editing.

e. Set F1 to the frequency indicated on the Test

Record.

2. Set up the Spectrum Analyzer as follows:

a. Span: 5 kHz/div

b. CF: Set to the 682XXB/683XXB frequency

value.

c. RBW: 1 kHz

d. Video Filter Wide: On

3. Adjust the Spectrum Analyzer Peaking control

for maximum signal level, then adjust the Refer-
ence Level Control to place the signal at the top
of the screen graticule.

4. Change the Spectrum Analyzer CF to each of the

harmonic frequencies listed on the Test Record
and record the signal levels. Refer to Table 3-4 for
the specified harmonic signal level limits.

5. Repeat steps 1 through 4 for each of the 682XXB/

683XXB CW carrier and harmonic frequencies
listed on the Test Record. Record the harmonic
signal levels on the Test Record.

682XXB/683XXB MM

3-19

PERFORMANCE

HARMONIC TEST: RF OUTPUT

VERIFICATION

SIGNALS FROM 2 TO 20 GHz

Harmonic and Harmonic Related:

500 MHz to

£

2.2 GHz (68XX5B):

<–50 dBc

10 MHz to

£

50 MHz:

<–30 dBc

>50 MHz to

£

2 GHz:

<–40 dBc

>2 GHz (2.2 GHz for 68XX5B)

to

£

20 GHz:

<–60 dBc

>20 GHz to

£

40 GHz:

<–40 dBc

Harmonic and Harmonic Related (Models having a
high-end frequency of >40 GHz and units with Op-
tion 15B at maximum specified leveled output
power):

500 Mhz to

£

2.2 GHz (68XX5B):

<–50 dBc

10 MHz to

£

50 MHz:

<–30 dBc

>50 MHz to

£

2 GHz:

<–40 dBc

>2 GHz (2.2 GHz for 68XX5B)

to

£

20 GHz:

<–50 dBc

>20 GHz to

£

40 GHz:

<–40 dBc

50 GHz units: >40 GHz to

£

50 GHz:

<–40 dBc

60 GHz units: >40 GHz to

£

60 GHz:

<–30 dBc

65 GHz units: >40 GHz to

£

45 GHz:

<–25 dBc

>45 GHz to

£

65 GHz:

<–30 dBc

Non-Harmonics:

500 MHz to

£

2.2 GHz (68XX5B):

<–50 dBc

10 MHz to

£

2 GHz:

<–40 dBc

>2 GHz (2.2 GHz for 68XX5B)

to

£

65 GHz:

<–60 dBc

Table 3-4.

Spurious Signals Specifications

Advertising
This manual is related to the following products: