3-8 Spurious Signals Test: RF Output Signals <2 GHz

Test Procedure

Measured Value

Upper Limit

Set F1 to 10 MHz
Record the presence of the worst case harmonic . . . . . . . . . .

Record the presence of the worst case non-harmonic . . . . . . .

_______________dBc

_______________dBc

–30 dBc

–40 dBc

Set F1 to 20 MHz
Record the presence of the worst case harmonic . . . . . . . . . .

Record the presence of the worst case non-harmonic . . . . . . .

_______________dBc

_______________dBc

–30 dBc

–40 dBc

Set F1 to 30 MHz
Record the presence of the worst case harmonic . . . . . . . . . .

Record the presence of the worst case non-harmonic . . . . . . .

_______________dBc

_______________dBc

–30 dBc

–40 dBc

Set F1 to 40 MHz
Record the presence of the worst case harmonic . . . . . . . . . .

Record the presence of the worst case non-harmonic . . . . . . .

_______________dBc

_______________dBc

–30 dBc

–40 dBc

Set F1 to 350 MHz
Record the presence of the worst case harmonic . . . . . . . . . .

Record the presence of the worst case non-harmonic . . . . . . .

_______________dBc

_______________dBc

–40 dBc

–40 dBc

Set F1 to 1.6 GHz
Record the presence of the worst case non-harmonic . . . . . . .

_______________dBc

–40 dBc

Set F1 to 1.6 GHz
Record the level of the harmonics of the 1.6 GHz carrier:

3.2 GHz (2nd harmonic) . . . . . . . . . . . . . . . . . . . . . .

4.8 GHz (3rd harmonic) . . . . . . . . . . . . . . . . . . . . . .

_______________dBc

_______________dBc

–40 dBc

–40 dBc

TEST

MODEL

RECORD

68247B/68347B

Advertising