Troubleshooting, Troubleshooting tables – Anritsu 682XXB User Manual

Page 205

Advertising
background image

682XXB/683XXB MM

5-39

YIG-tuned Oscillator

Error 124 Full Band Unlocked and Unleveled
Error 125 8.4-20 GHz Unlocked and Unleveled
Error 126 2-8.4 GHz Unlocked and Unleveled

Description: These error messages indicate either a failure of one or
both of the oscillators in the 2-20 GHz YIG-tuned oscillator assembly.

Step 1.

Connect a 562 Scalar Network Analyzer to the 682XXB/
683XXB as follows:

a.

Connect the 682XXB/683XXB

AUX I/O

to the 562

AUX

I/O

.

b.

Connect the 562

DEDICATED GPIB

to the 682XXB/

683XXB

IEEE-488 GPIB

.

c.

Connect the RF Detector to the 562 Channel A Input.

Step 2.

Set up the 682XXB/683XXB as follows:

a.

682XXB Setup:
CW/SWEEP SELECT: Step
F1: 2.000 GHz
F2: 20.000 GHz
Number of Steps: 400

683XXB Setup:
CW/SWEEP SELECT: Analog
F1: 2.000 GHz
F2: 20.000 GHz
Sweep Time: 0.100 Sec

Step 3.

Set up the 562 Scalar Network Analyzer as follows:

a.

Press SYSTEM MENU key.

b.

From System Menu display, select RESET.

c.

Press CHANNEL 2 DISPLAY: OFF.

d.

Press CHANNEL 1 DISPLAY: ON.

e.

Press CHANNEL 1 MENU key.

f.

From the Channel 1 Menu display, select POWER.

Step 4.

Using the scalar network analyzer, measure the RF output
directly at the YIG-tuned oscillator’s output connector. The
amplitude of the RF signal should be >4 dBm throughout
the full sweep.

Table 5-18.

Error Messages 124, 125, and 126 (1 of 2)

TROUBLESHOOTING

TROUBLESHOOTING

TABLES