Fm calibration calibration – Anritsu 682XXB User Manual

Page 154

Advertising
background image

Models 682XXB with Firmware Version 2.18 and above
Models 683XXB with Firmwave Version 2.20 and above
Models 682X5B/683X5B with Firmware Version 1.07 and above
This procedure provides the steps necessary to perform FM calibra-
tion. This consists of calibrating the FM Meter circuit and the FM
Variable Gain Control DAC. The FM Variable Gain Control DAC is
calibrated for input sensitivities in Locked, Locked Low-Noise, Un-
locked Narrow, and Unlocked Wide FM modes.

Equipment
Setup

Connect the equipment, shown in Figure 4-13, as
follows:

1. Interface the PC to the 682XXB/683XXB by per-

forming the initial setup procedure, pages 4-7 to
4-12.

2. Connect the 682XXB/683XXB rear panel

10 MHz

REF OUT

to the Spectrum Analyzer External Ref-

erence input.

3. Connect the Function Generator Output to the

BNC tee. Connect one leg of the tee to the
682XXB/683XXB front panel

FM IN

. Connect the

other leg of the tee to the DMM input.

4-44

682XXB/683XXB MM

FM

CALIBRATION

CALIBRATION

Serial

I/O

682XXB / 683XXB

COM1

or

COM2

IBM-Compatible PC

Function

Generator

10 MHz

REF OUT

EXT REF

INPUT

RF

IN

Spectrum

Analyzer

OUTPUT

FM

IN

RF

OUTPUT

BNC
Tee

INPUT

DMM

Figure 4-13.

Equipment Setup for FM Calibration

Advertising