This procedure provides the steps necessary to perform ALC Slope
calibration. The ALC Slope is calibrated to adjust for decreasing out-
put power-vs-output frequency in full band analog sweep.

Equipment
Setup

Connect the equipment, shown in Figure 4-7, as fol-
lows:

1. Interface the PC to the 683XXB by performing

the initial setup procedure, pages 4-7 to 4-12.

2. Using the Auxiliary I/O cable, connect the

683XXB rear panel

AUX I/O

connector to the 562

Network Analyzer AUX I/O connector.

3. Using the GPIB cable, connect the 562 Network

Analyzer DEDICATED GPIB connector to the
683XXB

IEEE-488 GPIB

connector.

4. Connect the RF Detector to the 562 Network

Analyzer Channel A Input connector.

5. Connect the 683XXB

RF OUTPUT

connector to

the RF Detector via a 10 dB Attenuator.

682XXB/683XXB MM

4-27

ALC SLOPE

CALIBRATION

CALIBRATION

Serial

I/O

683XXB Signal Generator

COM1

or

COM2

IBM-Compatible PC

562 Network Analyzer

10 dB

Attenuator

GPIB

AUX

I/O

RF Detector

Dedicated

GPIB

RF

Output

A

AUX

I/O

Figure 4-7.

Equipment Setup for ALC Slope Calibration