Coarse Loop/
YIG Loop Test
Procedure

The following procedure tests both the coarse loop
and YIG loop by stepping the signal generator
through its full frequency range in 1 GHz steps and
measuring the RF output at each step.

1. Set up the 682XXB/683XXB as follows:

a. Reset the instrument by pressing

SYSTEM

,

then

Reset

. Upon reset, the CW Menu is dis-

played.

b. Press

Edit F1

to open the current frequency

parameter for editing.

c. Set F1 to the first test frequency indicated on

the Test Record for the model being tested.

2. Record the Frequency Counter reading on the

Test Record. The Frequency Counter reading
must be within

±100 Hz of the displayed 682XXB/

683XXB frequency to accurately complete this
test.

NOTE

The Frequency Counter reading is typi-
cally within

±1 Hz because the instruments

use a common time base. Differences of a
few Hertz can be caused by noise or counter
limitations. Differences of

³±100 Hz indi-

cate a frequency synthesis problem.

3. On the 682XXB/683XXB, use the cursor control

key (diamond-shaped key) to increment F1 to the
next test frequency on the Test Record. Record
the Frequency Counter reading on the Test Re-
cord.

4. Repeat step 3 until all frequencies listed on the

Test Record have been recorded.