Performance phase verification modulation tests – Anritsu 682XXB User Manual

Page 110

Advertising
background image

Narrow

FM Mode

1. Set up the 682XXB/683XXB as follows:

a. Reset the instrument by pressing

SYSTEM

,

then

Reset

. Upon reset, the CW Menu is dis-

played.

b. Press

Edit F1

to open the current frequency

parameter for editing.

c. Set F1 to 5.0 GHz.

2. Set up the Spectrum Analyzer as follows:

a. CF: 5.0 GHz

b. Span/Div: 100 kHz (1 MHz across the display)

3. On the 682XXB/683XXB, make the following set-

tings:

a. Press

MODULATION

, then

FM . At the re-

sulting

FM Status menu, press

More

to go to

the additional

FM Status menu.

b. At the additional

FM Status menu, press

Narrow

, then press

Previous Menu

to return

to the main

FM Status menu.

c. At the

FM Status menu, press

Edit Dev.

and

set the deviation to 2.40 rad.

d. Press

Edit Rate

and set the rate to 99.8 kHz.

e. Press

On/Off

to turn

FM on.

4. On the 682XXB/683XXB, use the cursor control

key to place the cursor under the last 0 in the de-
viation setting (2.40 rad). While observing the
first Bessel null (Figure 3-15), increase/decrease
the

FM deviation with the cursor control key to

reduce the carrier as low as possible.

5. Record the

FM deviation, displayed on the

682XXB/683XXB, on the Test Record.

6. Compute the

FM accuracy with the following for-

mula:

Accuracy in

M Deviation in rad

(

%)

(

)

.

=

ж
и

з

ц
ш

ч ´

F

240

100

7. The calculated result should be between 93.7%

and 106.3%. Record this result on the Test Re-
cord.

3-50

682XXB/683XXB MM

PERFORMANCE

PHASE

VERIFICATION

MODULATION TESTS

NOTE

You may need to adjust the RBW
setting on the Spectrum Analyzer in
order to see the >–48 dBc null.

Advertising
This manual is related to the following products: