5 connector and key label notation 4-4, Following subassembly replacement, 4-4, Calibration following subassembly – Anritsu 682XXB User Manual

Page 114: Replacement -4, Connector and key label notation -4

Advertising
background image

4-4

CALIBRATION
FOLLOWING
SUBASSEMBLY
REPLACEMENT

Table 4-2 (page 4-6) lists the calibration that should be performed fol-
lowing the replacement of 682XXB/683XXB subassemblies or RF com-
ponents.

4-5

CONNECTOR AND KEY
LABEL NOTATION

The calibration procedures include many references to equipment in-
terconnections and control settings. For all 682XXB/683XXB refer-
ences, specific labels are used to denote the appropriate menu key,
data entry key, data entry control, or connector (such as

CW/SWEEP

SELECT

or

RF OUTPUT

). Most references to supporting test equip-

ment use general labels for commonly used controls and connections
(such as Span or RF Input). In some cases, a specific label is used that
is a particular feature of the test equipment listed in Table 4-1.

4-4

682XXB/683XXB MM

CALIBRATION FOLLOWING

CALIBRATION

SUBASSEMBLY REPLACEMENT

INSTRUMENT

CRITICAL

SPECIFICATION

RECOMMENDED

MANUFACTURER/MODEL

PROCEDURE

NUMBER

Frequency Counter

Frequency Range: 1 to 20 GHz
Input Impedance: 50

W

Resolution: 1 Hz

EIP Microwave, Inc. Model 578B

4-7

Spectrum Analyzer

Frequency Range: 1 to 20 GHz
Resolution Bandwidth: 10 Hz

Tektronix, Model 2794

4-13, 4-14

Power Meter,
with
Power Sensor

Power Range:–30 to +20 dBm

(1

m

W to 100mW)

Hewlett-Packard Model 437B, with
Power Sensor:

HP 8487A (0.01 to 50 GHz)

4-12

Power Meter
with
Power Sensor

Power Range:–30 to +20 dBm

(1

m

W to 100mW)

Anritsu ML4803A, with
Power Sensor:

MP716A4 (50 to 75 GHz

4-12

Frequency Standard

Frequency: 10 MHz
Accuracy: 1

´

10

–10

parts/day

Spectracom Corp., Model 8161

4-7

Function Generator

Output Voltage: 2 volts peak-to-peak
Functions: 0.1 Hz to 100 kHz sine and
square waveforms

Hewlett-Packard, Model 8116A

4-12, 4-13,

4-14

Function Generator

Output Voltage: 2 volts peak-to-peak
Functions: 0.1 Hz to 100 kHz sine and
square waveforms

Hewlett-Packard, Model 33120A

4-12. 4-13.

4-14

Digital Multimeter

Resolution: 4-1/2 digits (to 20V)
DC Accuracy: 0.002% +2 counts
DC Input Impedance: 10 M

W

AC Accuracy: 0.07% +100 counts

(to 20 kHz)

AC Input Impedance: 1 M

W

John Fluke, Inc.,Model 8840A, with
Option 8840A-09K (True RMS AC)

4-12, 4-13,

4-14

Table 4-1.

Recommended Test Equipment for Calibration Procedures (1 of 2)