Anritsu 682XXB User Manual

Page 88

Advertising
background image

c. With the Step Sweep Menu displayed, press

the main menu key

FREQUENCY

CONTROL

The Sweep Frequency Control Menu is then
displayed.

d. Press

Full

to select a full range frequency

sweep.

e. Press

Edit L1

to open the current power level

parameter for editing.

f. Set L1 to the power level indicated on the test

record.

g. Now, return to the Step Sweep Menu by press-

ing the main menu key

CW/SWEEP

SELECT

h. At the Step Sweep Menu, press

Sweep Ramp

to go to the Step Sweep Ramp Menu.

i. At this menu, press

Dwell Time

to open the

dwell time-per-step parameter for editing.

j. Set the dwell time to 1 second.

NOTE

Monitor the 682XXB/683XXB’s Horizontal
Output on the Oscilloscope to determine
sweep start and stop.

2. As the 682XXB/683XXB steps through the full

frequency range, measure the maximum and
minimum Power Meter readings and record the
values on the Test Record. Verify that the varia-
tion (difference between the maximum and mini-
mum readings) does not exceed the value noted
on the Test Record.

NOTE

This concludes power level testing for se-
ries 682XXB signal generators. For series
683XXB signal generators, continue on to
step 3 to test power level flatness in the
analog sweep mode.

3-28

682XXB/683XXB MM

PERFORMANCE

POWER LEVEL ACCURACY

VERIFICATION

AND FLATNESS TESTS