Error 147 Internal FM circuitry failed

Description: Indicates a failure of the internal frequency modulation
function. The 682XXB/683XXB may or may not provide frequency
modulation of the RF output signal using modulating signals from an
external source.

Step 1.

Set up the 682XXB/683XXB as follows:

a.

Press

SYSTEM

, then

Reset

.

b.

Press

MODULATION

, then

FM

to go to the Internal FM

Status display.

c.

Press

On/Off

to turn internal frequency modulation on.

Step 2.

Using an oscilloscope, verify the presence of a 10 volt peak to
peak sine wave signal with a period of 100

ms at the rear

panel

FM OUT

connector.

G

If present, replace the A11 PCB.

G

If not present, replace the A8 PCB.

Table 5-26.

Error Message 147

Error 148 Pulse 40 MHz reference circuitry failed.

Description: Indicates a failure of the pulse generator 40 MHz oscilla-
tor circuitry. The pulse generator may still function; however, the
40 MHz oscillator is not phase locked to the 10 MHz reference time-
base. The pulse modulation function may or may not operate. Error 106
(Power Supply not Phase-locked) may also be displayed.

Step 1.

Disconnect cable W108 at A6J1.

Step 2.

Using an oscilloscope, verify the presence of a 10 MHz at the
end of cable W108.

G

If present, replace the A6 PCB.

G

If not present, go to step 3.

Step 3.

Reconnect cable W108 to A6J1 and disconnect it at A5J4.

Step 4.

Using the oscilloscope, verify the presence of a 10 MHz TTL
signal at A5J4.

G

If present, replace the cable W108.

G

If not present, go to step 5.