Fm calibration calibration – Anritsu 682XXB User Manual

Page 160

Advertising
background image

c. Set up the Function Generator for a 99.8 kHz

sine wave with an output level of 0.707 volts
RMS (2 volts peak to peak). Use a frequency
counter to verify the output frequency of your
function generator is set to

±1%.

d. On the Spectrum Analyzer, set the Span/Div to

50 kHz per division.

e. On the computer keyboard, use the ‘, 1, 2. and

3 keys to increment and the 7, 8, 9, and 0 keys
to decrement the value of the DAC’s setting.
Start calibration by pressing an increment key.

f. While observing the first Bessel null (Figure

4-14) on the Spectum Analyzer display, adjust
the value of the DAC’s setting to reduce the
carrier level as low as possible.

g. When finished setting the DAC, press

Q

on the

keyboard to exit the calibration routine.

h. Record step completion on the Test Record.

8. Store the calibration data as follows:

a. At the

$

prompt, type:

calterm 787

and

press <ENTER>. (The

$

prompt will appear on

the screen whe the calibration data has been
stored.)

b. Record step completion on the Test Record.

4-50

682XXB/683XXB MM

FM

CALIBRATION

CALIBRATION

NOTE

You may need to adjust the RBW
setting on the Spectrum Analyzer in
order to see the >–50 dBc null.

CAUTION

When saving calibration data, turn-
ing off the instrument before the

$

prompt returns to the screen can
cause all stored data to be lost.