Troubleshooting, Troubleshooting tables – Anritsu 682XXB User Manual

Page 209

Advertising
background image

682XXB/683XXB MM

5-43

Step 7.

Connect a 562 Scalar Network Analyzer to the 682XXB/
683XXB as follows:

a.

Connect the 682XXB/683XXB

AUX I/O

to the 562

AUX

I/O

.

b.

Connect the 562

DEDICATED GPIB

to the 682XXB/

683XXB

IEEE-488 GPIB

.

c.

Connect the RF Detector to the 562 Channel A Input.

Step 8.

Set up the 562 Scalar Network Analyzer as follows:

a.

Press SYSTEM MENU key.

b.

From System Menu display, select RESET.

c.

Press CHANNEL 2 DISPLAY: OFF

d.

Press CHANNEL 1 DISPLAY: ON

e.

Press CHANNEL 1 MENU key.

f.

From the Channel 1 Menu display, select POWER.

Step 9.

Using the scalar network analyzer, measure the RF output
at J3 of the switched filter assembly. The amplitude of the
RF signal should be >17 dBm throughout the full sweep.

G

If the amplitude of the RF signal is correct, replace the
down converter assembly.

G

If there is no RF signal or if the amplitude of the RF sig-
nal is low, replace the switched filter assembly.

Table 5-20.

Error Messages 128, 129, 130, 131, 132, 133, 134, and 135 (3 of 8)

TROUBLESHOOTING

TROUBLESHOOTING