Troubleshooting, Troubleshooting tables – Anritsu 682XXB User Manual

Page 201

Advertising
background image

682XXB/683XXB MM

5-35

A11 FM PCB

Error 116 FM Loop Gain Check Failed

Description: The FM loop has failed or the loop gain is out of toler-
ance.

Step 1.

Perform a preliminary calibration. (Refer to chapter 4 for the
calibration procedure.)

Step 2.

Run self-test.

G

If error 116 is not displayed, the problem is cleared.

G

If error 116 is still displayed, go to step 3.

Step 3.

Replace the A11 PCB and run self-test again.

G

If error 116 is not displayed, the problem is cleared.

G

If error 116 is displayed, contact your local

ANRITSU

serv-

ice center for assistance.

Table 5-15.

Error Message 116

TROUBLESHOOTING

TROUBLESHOOTING

TABLES

Advertising
This manual is related to the following products: