Troubleshooting, Troubleshooting tables – Anritsu 682XXB User Manual

Page 184

Advertising
background image

5-18

682XXB/683XXB MM

Internal DVM Tests

Error 100 DVM Ground Offset Failed, or
Error 101 DVM Positive 10V Reference, or
Error 102 DVM Negative 10V Reference

Description: The DVM circuitry, located on the A16 CPU Interface
PCB, is calibrated using the

±10 volts from the reference supplies on

the A12 Analog Instruction PCB. The error messages indicate a
calibration-related problem or a defective

±10 volt reference.

Step 1.

Perform a manual pre-calibration. (Refer to chapter 4 for the
calibration procedure.)

Step 2.

Run self-test.

G

If no error message is displayed, the problem is cleared.

G

If any of the error messages, 100, 101, and 102, are dis-
played, go to step 3.

Step 3.

Connect the negative lead of the DMM to A12TP1.

Step 4.

Measure the

±10V reference voltages at A12TP4 and

A12TP8. A12TP4 should be –10V

±0.036V; A12TP8 should

be +10V

±0.036V.

G

If the

±10V reference voltages are correct, go to step 5.

G

If incorrect, replace the A12 PCB.

NOTE

Even if the

±10V reference voltages are correct, there could

still be a malfunction of the DVM multiplexer on the A12
PCB or the DVM circuitry on the A16 CPU Interface PCB.

Step 5.

Replace the A12 PCB and run self-test again.

G

If no error message is displayed, the problem is cleared.

G

If any of the error messages, 100, 101, and 102, are dis-
played, go to step 6.

Step 6.

Replace the A16 PCB, then run self-test.

G

If no error message is displayed, the problem is cleared.

G

If any of the error messages, 100, 101, and 102, are dis-
played, contact your local

ANRITSU

service center for

assistance.

Table 5-5.

Error Messages 100, 101, and 102

TROUBLESHOOTING

TROUBLESHOOTING

TABLES