Test procedure (11 to 20 ghz) 3-20, Test procedure (11 to 20 ghz) -20 – Anritsu 682XXB User Manual

Page 80

Advertising
background image

Test
Procedure
(11 to 20 GHz)

The following procedure lets you measure the 11 to
20 GHz RF output harmonic levels to verify that
they meet specifications.

NOTE

Because an external mixer is required for
these measurements, the RF output flat-
ness of the signal generator is used to cor-
rect for; (1) variations caused by switching
from the fundamental input to the external
mixer input of the Spectrum Analyzer, and
(2) the flatness of the mixer.

1. Set up the 682XXB/683XXB as follows:

a. Reset the instrument by pressing

SYSTEM

,

then

Reset

. Upon reset the CW Menu is dis-

played.

b. Press

Edit F1

to open the current frequency

parameter for editing.

c. Set F1 to the frequency indicated on the Test

Record.

d. Press

Edit L1

to open the current power level

parameter for editing.

e. Set L1 to –30 dBm output power.

NOTE

If the 682XXB/683XXB is not fitted with
Option 2, install a 30 dB attenuator (AN-
RITSU 41KC-20 and 41KC-10 for

£40 GHz

models; 41V-20 and 41V-10 for >40 GHz
models) and set L1 to 0.0 dBm output
power.

2. Set up the Spectrum Analyzer as follows:

a. Span: 5 kHz/div

b. CF: Set to the 682XXB/683XXB frequency

value.

c. RBW: 1 kHz

d. Video Filter Wide: On

3. Adjust the Spectrum Analyzer Peaking control

for maximum signal, then adjust the Reference
Level control to place the signal at the top of the
screen graticule. It may be necessary to also ad-
just the 682XXB/683XXB output power level

3-20

682XXB/683XXB MM

PERFORMANCE

HARMONIC TEST: RF OUTPUT

VERIFICATION

SIGNALS FROM 2 TO 20 GHz

Advertising