Anritsu 682XXB User Manual

Page 261

Advertising
background image

682XXB/683XXB MM

A-23

3-9 Harmonic Test: RF Output Signals From 2.2 to 20 GHz (Continued)

Test Procedure (11 to 20 GHz)

Measured Value

Upper Limit

Set F1 to 12.4 GHz
Record the level of all harmonics of the 12.4 GHz carrier:

24.8 GHz (2nd harmonic) . . . . . . . . . . . . . . . . . . . . .

37.2 GHz (3rd harmonic) . . . . . . . . . . . . . . . . . . . . . .

_______________dBc

_______________dBc

–60 dBc*

–60 dBc*

Set F1 to 16.0 GHz
Record the level of all harmonics of the 16.0 GHz carrier:

32.0 GHz (2nd harmonic) . . . . . . . . . . . . . . . . . . . . .

_______________dBc

–60 dBc*

Set F1 to 20.0 GHz
Record the level of all harmonics of the 20.0 GHz carrier:

40.0 GHz (2nd harmonic) . . . . . . . . . . . . . . . . . . . . .

_______________dBc

–60 dBc*

* –50 dBc if Option 15B (High Power) installed.

TEST

MODEL

RECORD

68245B/68345B

Advertising
This manual is related to the following products: