3-4

682XXB/683XXB MM

INSTRUMENT

CRITCAL

SPECIFICATION

RECOMMENDED

MANUFACTURER/MODEL

TEST

NUMBER

Spectrum Analyzer,
with
External Mixers
and
Diplexer Assy

Frequency Resolution: 0.01 to 65 GHz
Resolution Bandwidth: 10 Hz

Tektronix, Model 2794, with
External Mixers:

WM780K (18 to 26.5 GHz)
WM780A (26.5 to 40 GHz)
WM780U (40 to 60 GHz)
WM780E (60 to 90 GHz)

and
Diplexer Assy: 015-0385-00

3-8, 3-9, 3-12,

3-13, 3-16,

3-17

Spectrum Analyzer

Frequency Range: 20 Hz to 40 MHz
Resolution Bandwidth: 3 Hz

Hewlett-Packard, Model 3585B

3-10

Frequency Counter
with
Cable Kit
and
External Mixer

Frequency Range: 0.01 to 65 GHz
Input Impedance: 50

Resolution: 1 Hz
Other: External Time Base Input

EIP Microwave, Inc. Models 538B,
548B, or 578B, with
Cable Kit: Option 590
and
External Mixer:

Option 91 (26.5 to 40 GHz)
Option 92 (40 to 60 GHz)
Option 93 (60 to 90 GHz)

3-7

Power Meter,
with
Power Sensor

Power Range: –30 to +20 dBm

(1

m

W to 100mW)

Hewlett-Packard Model 437B, with
Power Sensor:

HP 8487A (0.01 to 50 GHz)

3-11

Power Meter
with
Power Sensor

Power Range: –30 to +20 dBm

(1

m

W to 100mW)

ANRITSU ML4803A, with
Power Sensor:

MP716A4 (50 to 75 GHz)

3-11

Frequency Reference

Frequency: 10 MHz
Accuracy: 5 x 10

–12

parts/day

Absolute Time Corp., Model 300

3-6

Modulation Analyzer

Frequency Input: 10 MHz

(or the IF of the spectrum analyzer)

AM Depth: 0% to 90%
AM Modulation Rates: dc to 100 kHz
Filters: 20 kHz lowpass, 300 Hz highpass

Hewlett-Packard, Model 8901A

3-12

Oscilloscope

Bandwidth: DC to 150 MHz
Vertical Sensitivity: 2mV/division
Horizontal Sensitivity: 50 ns/division

Tektronix, Inc. Model TAS485

3-6, 3-11,

3-14, 3-15

Mixer

Frequency Range: 1 to 26 GHz

Macom Micro Electronics Div.
Model DMS1-26A

3-10

Adapter

K (male) to 2.4 mm (female)
Adapts the Power Sensor, HP 8487A, to
the 682XXB/683XXB RF OUTPUT con-
nector (

£

40 GHz models)

Hewlett-Packard
Part Number: HP 11904D

3-11

Adapter

Adapts the MP716A4 Power Sensor to
the ML4803A Power Meter

ANRITSU MA4002B

3-11

Table 3-1.

Recommended Test Equipment for Performance Verification Tests (1 of 2)

PERFORMANCE

RECOMMENDED

VERIFICATION

TEST EQUIPMENT